Medical air should be sampled at which type of valve sample port?

½" NPT

¼" NPS

Medical air should be sampled at a ¼" NPS valve sample port because this size allows for an optimal flow rate and pressure for accurate sampling of the medical air supply. The ¼" NPS (National Pipe Straight) configuration is standard for gas sampling ports in medical systems, ensuring compatibility with sampling equipment and regulatory standards.

This size also balances the need for obtaining a representative sample while minimizing the risk of contamination or pressure drops that might occur with larger ports. In medical gas systems, maintaining the integrity of the sample is crucial, and the ¼" NPS size helps achieve this by allowing for effective monitoring of gas quality and composition.

In contrast, other options—like the ½" NPT or 1" BSP—are generally used for different applications, such as larger gas delivery systems where higher flow rates are necessary, or they might not be standard for sampling in clinical settings. The ¾" NPT option, while also a standard thread size, does not provide the optimal sampling conditions that the ¼" NPS does. Thus, the use of the ¼" NPS port is specifically tailored for ensuring accurate and safe sampling processes in medical air systems.